TPS 500

The Hot Disk® TPS 500 is an economical step-in TPS instrument for fast and accurate measurement of Thermal Conductivity, Thermal Diffusivity, Thermal Effusivity and Specific Heat Capacity in everyday laboratory and QC use.

Overview

The Hot Disk® TPS 500 swiftly and accurately measures the Thermal Conductivity, Thermal Diffusivity, Thermal Effusivity and Specific Heat Capacity of a wide range of materials. It is therefore a favorable step-in thermal conductivity meter, and highly suitable for QC testing. The TPS 500 can handle bulk sample sizes down to a thickness of a few millimeters. The TPS 500 Thermal Conductivity range is from 0.03 to 100 W/m/K, and its temperature range from -100°C to 200°C. It is partly based on ISO 22007-2* and ASTM E3088-25*, and is CE marked.

– With revised electronics.

For the economical lab

The accessible Hot Disk® TPS 500 Thermal Constants Analyser quickly and accurately measures Thermal Conductivity, Thermal Diffusivity, Thermal Effusivity, and Specific Heat Capacity of a wide range of materials and is therefore an excellent step-in thermal conductivity meter for the economical laboratory or QC testing set-up.

The TPS 500 instrument condenses our patented and now ISO and ASTM standardised TPS technology with our user-friendly test and analysis software, resulting in a smart, simple-to-use thermal conductivity meter.

The TPS 500 measures the Thermal Transport Properties of solids, powders, pastes, gels and liquids over a temperature range of -100°C to 200°C. The instrument encompasses similar accuracy and sample size flexibility as the benchmark TPS 2500 S and workhorse TPS 1500 (the latter two instruments are designed according to ISO 22007-2 and ASTM E3088-25).

TPS 500 Specifications1

Sample Types Bulks, fine granules, powders, pastes, creams, gels, liquids, and foams.
Evaluation
Anisotropy No.
One-Dimensional No.
Sample Dimensions
Smallest 3 mm thick x 13 mm wide (circle or square).
Largest Unlimited.
Sample Temperature Range -100°C to 200°C.
Core Instrument RT.
With Optional TCU Cryostat: -100°C to RT.
Low-to-High Temperature Chamber: -60°C to 200°C.
Bath Circulator: -40°C to 200°C.
Convection Oven: RT to 200°C.
Muffle Furnace: RT to 200°C.
Tubular Furnace: RT to 200°C.
Measurement Time2 2.5 to 2560 seconds (depending on sample material and sensor dimensions).
Measurement Range
Thermal Conductivity 0.03 to 100 W/m/K.
Thermal Diffusivity 0.02 to 50 mm²/s.
Thermal Effusivity 40 to 20000 W√s/m²/K.
Specific Heat Capacity 0.1 to 4.5 MJ/m³/K.
Measurement Accuracy
Thermal Conductivity Better than 5%.
Thermal Diffusivity Better than 10%.
Measurement Reproducibility3
Thermal Conductivity Typically better than 2%.
Thermal Diffusivity Typically better than 10% (Kapton-insulated Hot Disk® sensor model 5501).
Measurement Repeatability4
Thermal Conductivity Typically 0.32% (Stainless Steel bulk sample, Kapton-insulated Hot Disk® sensor model 5501, and 22°C sample temperature).
Thermal Diffusivity Typically 1.3% (Stainless Steel bulk sample, Kapton-insulated Hot Disk® sensor model 5501, and 22°C sample temperature).
Measurement Sensitivity5
Temperature Typically 610 μK (Stainless Steel sample, Kapton-insulated Hot Disk® sensor model 5501, and 22°C sample temperature).

General

Meets ISO Standard 22007-2 Yes, with revised electronics.
Meets ASTM Standard E3088-25 Yes, with revised electronics.
Measurement Modules
Included Isotropic and Single-Sided Testing.
Optional Automation.
Sensor Models
Hot Disk® sensor – Kapton-insulated Models 5465 and 5501.
Hot Disk® sensor – Teflon-insulated Models 5465 and 5501.
Instrument Environment
Operating 15°C to 30°C
Storage -20°C to 45°C
Instrument Line Power
Supply Voltage 100 – 240 VAC, 50 – 60 Hz.
Power Consumption 18 W typical, 190 W maximum.
Instrument Warm-Up 2 hours to rated measurement performance.
Instrument Input and Output
Rear Panel USB type B connector (female) for connectivity.
RS232 connector (male) for optional TCU.
Front Panel LEMO 8 pin connector (female) for Hot Disk® sensor.
LEMO 4 pin connector (female) for optional PT100 temperature sensor.
Instrument Dimensions 190 mm high x 455 mm wide x 485 mm deep.
Instrument Weight 11 kg.
Instrument CE Marking Yes.
  1. Hot Disk AB reserves the right to make changes without prior notice, whether due to misprints, improved hardware, or extended software capabilities.
  2. Time of transient temperature reading. An additional temperature settling time that depends on the sample material and TCU (if applied) is required to ensure isothermal condition before the reading.
  3. Comparing measurements performed by different instrument set-ups.
  4. Variation in consecutive measurements performed by a single instrument set-up. Note that this value depends on sample material, sensor model, and sample temperature.
  5. Standard deviation in fitting ∆Τ(τ)-function to the transient temperature reading. Note that this value depends on sample material, sensor model, sample temperature, and can vary considerably.

Included with the TPS 500 instrument

  • 2 x Measurement Modules: Isotropic and Single-Sided Testing.
  • 2 x Kapton-insulated Hot Disk® sensors: Models 5465 and 5501, with cable.
  • Hot Disk® Desktop App software (latest version)
  • 1 x Stainless Steel verification bulk sample.
  • 1 x Single-sided background insulation (single-sided testing).
  • 1 x Room temperature sample holder (protective hood optional).
  • Instructions manual.
  • Instrument test protocol.
  • Power supply cord and USB cable.
Material properties
Specific Heat Capacity Thermal Conductivity Thermal Diffusivity Thermal Effusivity